FREE AND TOTAL CHLORINE MEASUREMENT

To take colorimetric measurements follow next
steps:
• Turn the instrument on by pressing

ON/OFF.

Assure that the correct range is selected by
paying attention to the startup message or to
the measuring units.

• When dashes are displayed on the LCD, the

instrument is ready. The current time appears
on the secondary LCD, if selected in SETUP
menu, or “

F Cl” or “t Cl” if the time is not

displayed. The “

ZERO” tag will blink

suggesting that a zero measurement should be
done.

ZEROING THE INSTRUMENT

• Fill a clean, dry cuvette with 10 mL of sample,

up to the mark, taking care to handle the cuvette
by the top. Replace the cap.

• Wipe the cuvette thoroughly with a lint-free

cloth to remove any fingerprints, dirt or water
spots.

• Place the cuvette into the instrument. Align the

mark on the cuvettee with the sign on the
instrument top and close the lid.

Note: If you have a cuvette with orientation mark

place the cuvettee with the orientation mark
aligned with the sign on the instrument top.

• Press

ZERO/AVG

. The dashes will blink

on the primary LCD. If the zeroing procedure
was successful, the display will show “

-0.0-”.