SCPI Command Reference

78

IQS-2600B

Product-Specific Commands—Description

:SOURce[1..n]:POWer[:LEVel]

[:IMMediate][:AMPLitude]

Description

This command sets the source output power

level. This value can be set even if the source is

off.

At *RST, the power level is -8.0 dBm.

Syntax

:SOURce[1..n]:POWer[:LEVel][:IMMediate][:AM

PLitude]DBM]>|

MAXimum|MINimum|DEFault

Parameter(s)

OutputPower:
The program data syntax for is

defined as a element

followed by an optional

DATA> element. The allowed

PROGRAM DATA> element is DBM. The

special forms MINimum,

MAXimum and DEFault are accepted on input.

MINimum allows to set the instrument to the

smallest supported value.

MAXimum allows to set the instrument to the

greatest supported value.

DEFault allows the instrument to select a value

for the parameter.

This parameter is the new source output power

value in dBm.