Source[1..n]:wavelength:sweep: step[:width – EXFO IQS-2600B Tunable Laser Source for IQS Platforms User Manual

Page 122

background image

SCPI Command Reference

116

IQS-2600B

Product-Specific Commands—Description

:SOURce[1..n]:WAVelength:SWEep:

STEP[:WIDTh]

Description

This command allows you to specify the

wavelength step for the step-by-step sweep.

At *RST, this value is device-dependent.

Syntax

:SOURce[1..n]:WAVelength:SWEep:STEP[:WIDT

h]M|HZ]>|MAXim

um|MINimum|DEFault

Parameter(s)

StepWidth:
The program data syntax for is

defined as a element

followed by an optional

DATA> element. The allowed

PROGRAM DATA> elements are: M|HZ. The

special forms MINimum,

MAXimum and DEFault are accepted on input.

MINimum allows to set the instrument to the

smallest supported value.

MAXimum allows to set the instrument to the

greatest supported value.

DEFault allows the instrument to select a value

for the parameter.

The parameter is the new step

width, in meters, for the sweep program in

Step-by-Step mode.