SCPI Command Reference

Tunable Laser Source

101

Product-Specific Commands—Description

:SOURce[1..n]:WAVelength:SWEep:

STARt

Description

This command allows you to specify a start

wavelength for the sweep.

At *RST, the start wavelength is 1520 nm.

Syntax

:SOURce[1..n]:WAVelength:SWEep:STARt

>M|HZ]>|MAXimum

|MINimum|DEFault

Parameter(s)

StartWavelength:
The program data syntax for

is defined as a element

followed by an optional

DATA> element. The allowed

PROGRAM DATA> elements are: M|HZ. The

special forms MINimum,

MAXimum and DEFault are accepted on input.

MINimum allows to set the instrument to the

smallest supported value.

MAXimum allows to set the instrument to the

greatest supported value.