Source[1..n]:wavelength:sweep: start – EXFO IQS-2600B Tunable Laser Source for IQS Platforms User Manual

Page 108

background image

SCPI Command Reference

102

IQS-2600B

Product-Specific Commands—Description

DEFault allows the instrument to select a value

for the parameter.

The parameter is the

desired start spectrum value for the sweep. If you

do not specify a unit, the device will use the

current spectral units (meter or Hz). To select the

current spectral units, use the

UNIT[1..n]:SPECtrum command.

Example(s)

SOUR:WAV:SWE:STAR 0.00000155 (if current unit

is meters)

SOUR:WAV:SWE:STAR 1521.157NM

SOUR:WAV:SWE:STAR 192.000 THZ

Notes

This command cannot be performed during a

sweep.

See Also

SOURce[1..n]:WAVelength:SWEep:STARt?

SOURce[1..n]:WAVelength:SWEep:STOP?

UNIT[1..n]:SPECtrum

:SOURce[1..n]:WAVelength:SWEep:

STARt