Calibration[1..n]:wavelength – EXFO IQS-2600B Tunable Laser Source for IQS Platforms User Manual

Page 80

background image

SCPI Command Reference

74

IQS-2600B

Product-Specific Commands—Description

Example(s)

SOUR:POW:STATE ON

- Wait for the power on operation to be complete.

SOUR:WAV:CW 1560E-009

- Wait for one hour.

- To recalibrate the complete wavelength range,

EXFO recommends using 1560.000 nm as the

reference.

- In this example, the output wavelength

measurement is taken. Supposing that the

measurement wavelength is 1560.023, then:

CAL:WAV 1560.023E-009

Notes

This function cannot be performed while the

module is sweeping.

The difference between the current wavelength

and the measured wavelength cannot be greater

then ± 0.200 nm.

EXFO recommends to calibrate the module at

1560 nm.

:CALibration[1..n]:WAVelength