beautypg.com

Source[1..n]:power:alc[:state – EXFO IQS-2600B Tunable Laser Source for IQS Platforms User Manual

Page 82

background image

SCPI Command Reference

76

IQS-2600B

Product-Specific Commands—Description

:SOURce[1..n]:POWer:ALC[:STATe]

Description

This command activates the laser output

Automatic Leveling Control (ALC). The source is

either in ALC mode or Maximum Power mode.

At *RST, the mode is ALC.

Syntax

:SOURce[1..n]:POWer:ALC[:STATe]

e>

Parameter(s)

State:
The program data syntax for is defined

as a element. The

special forms ON and OFF are accepted

on input for increased readability. ON

corresponds to 1 and OFF corresponds to 0.

The parameter is the new state of the

ALC.

ON or 1, the power automatic leveling control is

active.

OFF or 0, the power automatic leveling is not
active (maximum power).

Example(s)

SOUR:POW:ALC ON

Notes

This function cannot be performed while the

module is sweeping.

This command is for modules equipped with the

PMF output only.

See Also

SOURce[1..n]:POWer:ALC[:STATe]?