SCPI Command Reference

96

IQS-2600B

Product-Specific Commands—Description

:SOURce[1..n]:WAVelength:SWEep:

DIRection

Description

This command controls the direction of the

sweep. If ONEWay is selected, the sweep is

carried out from STARt to STOP. If TWOWay is

selected, the sweep is carried out from STARt to

STOP and from STOP to STARt.

At *RST, the value of this function is TWOWay.

Syntax

:SOURce[1..n]:WAVelength:SWEep:DIRection<

wsp>ONEWay|TWOWay

Parameter(s)

Directional Way:
The program data syntax for the first parameter is

defined as a

element. The allowed

DATA> elements for this parameter are:

ONEWay|TWOWay.

The parameter is indicating whether or not the
sweep will be performed in both directions:

ONEWay, the sweep is performed in one

direction, from start to stop.

TWOWay, the sweep is performed in two

directions, from start to stop to start.

Example(s)

SOUR:WAV:SWE:DIR BOTH

Notes

This function cannot be performed while the

module is sweeping.

See Also

SOURce[1..n]:WAVelength:SWEep:DIRection?