beautypg.com

Mportant – EXFO IQS-12002B Calibration System User Manual

Page 92

background image

Configuring the FTB-7000D OTDR on the IQS-12002B

88

IQS-12002B

Creating the GP-121723: 50-m Launch Fiber as a Calibration Instrument

4. From the View Standard Parameters window, click Edit to enter the

length of the reference standard given on the acceptance test plan for
all requested wavelengths (e.g., 56.1 m).

5. Click Close.

I

MPORTANT

Ensure that the Length Tolerance parameter is set to 5 m.