beautypg.com

EXFO IQS-12002B Calibration System User Manual

Page 100

background image

Configuring the FTB-7000D OTDR on the IQS-12002B

96

IQS-12002B

Creating the GP-120963: Singlemode Dead-Zone Calibration Instrument

1c. Enter the inventory number and comments if necessary.

Note: When the basic model of the calibration instrument is configured to appear

in the Calibration Certificate produced by the IQS-12002B, the information
that you enter in the Inventory number text box will appear in the section
entitled Standards used to establish traceability.

1d. Select Calibrated (no recalibration required).

2. Click OK.

3. From the Add a Calibration Event window, enter the date as written

on the instrument calibration certificate, and click OK.

4. Use the length of the reference standard given on the acceptance test

and add 3 meters to it (new length = 3 m + given length).