Introducing the IQS-12002B Calibration System

2

IQS-12002B

Some units (DUT) cannot be adjusted using the IQS-12002B. Depending on
the DUT model, adjustment will or will not be possible. When adjustment
is not possible, you can only perform calibration and verification of the unit.
If adjustment is possible, you will be asked to perform calibration and
verification of the DUT before and after adjustment.

The provided application helps you to comply to the ISO/IEC 17025
standard with the use of the following features:

³

The application keeps (in the database) the complete history of the
operations performed by a user on a DUT.

³

You can report all DUTs that have been verified or adjusted using a
specific standard.

³

For some procedures, the system allows taking uncertainties into
account when judging on compliance.