beautypg.com

Finding information on the exfo web site – EXFO IQS-12002B Calibration System User Manual

Page 48

background image

Troubleshooting

44

IQS-12002B

Finding Information on the EXFO Web Site

Finding Information on the EXFO Web Site

The EXFO Web site provides answers to frequently asked questions (FAQs)
regarding the use of your IQS-12002B Calibration System.

To access FAQs:
1.
Type

http://www.exfo.com

in your Internet browser.

2. Click on the Support tab.

3. Click on FAQs and follow the on-screen instructions. You will be given

a list of questions pertaining to your subject.

The EXFO Web site also provides the product’s most recent technical
specifications.

Error message

“Catastrophic failure”

Incorrect connections

for the FOT-10A or

FOT-10A unit.

Disconnect the unit, close all

applications, and restart the system.

Make sure the connection procedure is

properly done.

Calibration has a Fail

in Distance result.

The automatic

positioning of events

for an FTB-7000B-B

OTDR is incorrect.

Positioning must be done manually. Set

marker A right before the reflective

event at 50 m and marker B right before

the end of fiber event.

Interruption of the

dead zone calibration

The automatic

positioning of the

events is incorrect.

Make sure that the standard distance

information include the 3-m patchcord

as described in Configuring the
FTB-7000D OTDR on the IQS-12002B
on

page 79.

Impossible to print

information on a

specific DUT.

This DUT has never

been tested.

Test the DUT and try again.

Problem

Possible cause

Solution