beautypg.com

B configuring the ftb-7000d otdr on the iqs-12002b – EXFO IQS-12002B Calibration System User Manual

Page 83

background image

Calibration System

79

B Configuring the FTB-7000D

OTDR on the IQS-12002B

Test equipment is configured at the factory before shipping. The
procedures presented in this document address the case when the
IQS-12002B detects a new instrument or when a particular instrument has
not been configured.

Note: The given instructions are for the singlemode FTB-7000D OTDR; however,

you can use the same procedures for the multimode model (lengths,
tolerances, and uncertainties can differ).

The following table shows the list of calibration instruments and standards
for the singlemode and the multimode OTDRs.

Calibration Instruments and Standards

for Singlemode OTDR

Calibration Instruments and Standards

for Multimode OTDR

GP-121311: 4.4-km fiber spool

GP-121492: 1-km multimode fiber spool

GP-121317: 17.6-km fiber spool

GP-121490: 4-km multimode fiber spool

IQS-3100BW Variable Attenuator

IQS-3100B-D Variable Attenuator

GP-121723: 50-m launch fiber

Useless for multimode calibration

GP-121392: distance reference standard

GP-121489: multimode distance reference

standard

GP-121310: singlemode front-connector

position (for UPC connectors)

GP-121487: multimode front-connector

position (for UPC connectors)

GP-120963: singlemode dead zone

GP-121200: multimode dead zone

GP-121310: singlemode front-connector

position (for APC connectors)

GP-121487: multimode front-connector

position (for APC connectors)