A technical specifications, Mportant, Multifunction loss tester – EXFO FTB-3930 MultiTest User Manual
Page 93: Technical specifications
Multifunction Loss Tester
87
A Technical Specifications
I
MPORTANT
The following technical specifications can change without notice.
The information presented in this section is provided as a reference
only. To obtain this product’s most recent technical specifications,
visit the EXFO Web site at www.exfo.com.
TECHNICAL SPECIFICATIONS
External Power Meter
FTB-3932
FTB-3932X
FTB-3933
Detector type
Ge
GeX
InGaAs
Measurement range (dBm)
10 to –70
26 to –55
6 to –73
Uncertainty
b, c
± 5 % ± 0.1 nW
± 5 % ± 3 nW
± 5 % ± 0.05 nW
Wavelength range (nm)
800 to 1650
800 to 1650
800 to 1650
Display resolution
b
(dB)
0.01
0.01
0.01
Calibrated wavelengths
40
42
40
Recommended recalibration
period (years)
3
3
3
Automatic offset nulling
d
Yes
Yes
Yes
Measurement-distance units
kilometers, meters, kilofeet, feet, miles
Sources
Standard
–4
–5
–12C (second port)
–12D (second port)
Wavelengths
e
(nm)
1310 ± 20
1550 ± 20
1310 ± 20
1550 ± 20
1625 ± 10
1310 ± 20
1490 ± 10
1550 ± 20
850 ± 25
1300 +50/–20
850 ± 25
1300 +50/–20
Emitter type
Laser
Laser
Laser
LED
LED
Minimum output power
e
(dBm)
–1/–1
–1/–4/–7
–1/–7/–4
–27/–27 (50/125 μm)
i
–21/–21 (62.5/125 μm)
i
Spectral width
f
(nm)
) 5/) 5
) 5/) 5/) 5
) 5/) 5/) 5
50/135
50/135
Stability
g
(8 hours) (dB)
± 0.05
± 0.05
± 0.05
± 0.05
± 0.05
FasTesT
Standard
–4
–5
–12C (second port)
–12D (second port)
Wavelengths (nm)
1310
1550
1310
1550
1625
1310
1490
1550
850
1300
850
1300
Loss range
h
(dB)
60
56
56
40
46
Loss precision
i
(repeatability) (dB)
side-by-side
loopback
0.15
0.25
0.15
0.25
0.15
0.25
0.15
0.25
0.15
0.25
Length measurement
range (km)
200
200
200
5
5
Length measurement
uncertainty (typical)
j
± (10 m + 1 % x length)