Troubleshooting

78

FTB-3930

Solving Common Problems

Unable to establish F

AS

T

ES

T

communication.

Remote unit’s F

AS

T

ES

T

port is in use.

Remote unit is not

compatible.

Selected port on master

unit differs from actual

port to which fiber is

connected.

Compatibility set to
FOT-930 / FTB-3930

but remote unit is not

an FOT-930 or

FTB-3930.

Wait until F

AS

T

ES

T

completes, turn source

and ORL meter off or exit

Probe mode (FOT-930).

Make sure remote unit

isan FOT-930, FTB-3930,

FOT-920 or FTB-3920.

Set the port (SM or MM)

correctlyin Setup (F

AS

T

ES

T

tab) on master unit.

Set the compatibility to
FOT-920 / FTB-3920 in
Setup (

F

AS

T

ES

T tab) on

master unit.

During F

AS

T

ES

T, you get a

message saying that

loopback reference was

not performed.

Loopback reference was

selected, but not performed

(or reference is outdated).

Perform a loopback reference

before F

AS

T

ES

T measurement.

ORL values are inaccurate

for short fibers at 1310 nm

during a F

AS

T

ES

T.

Measure ORL manually.

Unable to communicate

with an FOT-920 using the

talk set.

Talk sets of the FOT-920 and

FOT-930/FTB-3930 are not

compatible.

Use text messaging instead.

Unable to establish

connection with

compatible talk set.

Probe mode is activated on

remote unit(FOT-930 only).

Deactivate Probe mode.

Problem

Possible Cause

Solution