Defining the list of favorite wavelengths, Favorite wavelengths – EXFO FTB-3930 MultiTest User Manual
Page 26

Measuring Power or Loss
20
FTB-3930
Defining the List of Favorite Wavelengths
Defining the List of Favorite Wavelengths
You must put the wavelengths you want to use on a list of favorite
wavelengths. Only wavelengths selected from this list are available for
measurements.
By default, the list contains 22 of the 40 calibrated wavelengths. It can
contain a maximum of 30 wavelengths.
Specifications are guaranteed for calibrated wavelengths only. For other
wavelengths, the unit will determine values based on the calibrated
wavelengths (3-point interpolation).
Note: The list must always contain at least one selected wavelength.
Detector
Type
Calibrated
Wavelengths (nm)
Default Favorite
Wavelengths (nm)
InGaAs
Ge
800, 820, 830, 840, 850, 860, 870,
880, 910, 980, 1270, 1280, 1290,
1300, 1310, 1320, 1330, 1340, 1390,
1450, 1460, 1470, 1480, 1490, 1500,
1510, 1520, 1530, 1540, 1550, 1560,
1570, 1580, 1590, 1600, 1610, 1620,
1630, 1640, 1650.
800, 840, 850,
860, 910, 980,
1280, 1300, 1310,
1320, 1450,
1470, 1480, 1490,
1510, 1520, 1530, 1540,
1550, 1560,1570, 1625.
GeX
All the above, plus 1370 and 1060.
Same as above.