Measuring Optical Return Loss

32

FTB-3930

Performing and Saving ORL Measurements

Performing and Saving ORL Measurements

You can define

ORL thresholds

(see Setting Pass/Fail Thresholds on

page 16) before or after measuring ORL. ORL values below thresholds are
displayed in red.

To measure ORL:
1.
If necessary, perform an

offset nulling

(see Nulling Electrical Offsets on

page 22).

2. From the main window, select the ORL Meter tab.

3. Scroll through the Wavelength list to select a singlemode wavelength.

4. Verify your patchcords and clean them properly (see Cleaning and

Connecting Optical Fibers on page 12).

Note: If the F

AS

T

ES

T SM port of your unit is equipped with an APC connector, use

EXFO’s optional ORL calibrated patchcord.