Referencing units for fastest – EXFO FTB-3930 MultiTest User Manual
Page 44

Performing Automated IL/ORL/Length Measurements (FasTesT)
38
FTB-3930
Referencing Units for FasTesT
Referencing Units for FasTesT
Referencing subtracts the loss caused by the test setup components from
the overall loss measured during F
AS
T
ES
T. The final result represents the
loss inserted by the system under test alone.
Two referencing methods are available:
Side-by-Side Method (Best)
Loopback Method
Description
Reference taken with both units
together using their F
AS
T
ES
T ports.
Slightly more accurate value than
loopback method.
Reference taken separately on
each unit (F
AS
T
ES
T port connected
to power meter port).
Location of units Must be at same location.
Can be at different locations.
Loss included in
F
AS
T
ES
T result
Loss due to system under test and
one connector mating.
Loss due to system under test and
the two connector matings.
Elements to
consider
Includes neither an ORL
reference nor an
. To obtain them,
use the ORL Meter pane (see
Setting ORL Zero Value on
page 30).
With multiple referencing, you
may coordinate an FTB-3930 with
up to 10 FOT-930 units.
When measuring ORL (F
AS
T
ES
T or
ORL meter), accounts for
connector loss and adjusts ORL
calibration accordingly.
Not recommended for short links.
System
under test
Connector
mating
Connector
mating
System
under test
Connector
mating