Referencing units for fastest – EXFO FTB-3930 MultiTest User Manual

Page 44

background image

Performing Automated IL/ORL/Length Measurements (FasTesT)

38

FTB-3930

Referencing Units for FasTesT

Referencing Units for FasTesT

Referencing subtracts the loss caused by the test setup components from
the overall loss measured during F

AS

T

ES

T. The final result represents the

loss inserted by the system under test alone.

Two referencing methods are available:

Side-by-Side Method (Best)

Loopback Method

Description

Reference taken with both units

together using their F

AS

T

ES

T ports.

Slightly more accurate value than

loopback method.

Reference taken separately on

each unit (F

AS

T

ES

T port connected

to power meter port).

Location of units Must be at same location.

Can be at different locations.

Loss included in
F

AS

T

ES

T result

Loss due to system under test and

one connector mating.

Loss due to system under test and

the two connector matings.

Elements to

consider

Includes neither an ORL

reference nor an

ORL zero

measurement

. To obtain them,

use the ORL Meter pane (see

Performing ORL Reference and

Setting ORL Zero Value on
page 30
).
With multiple referencing, you

may coordinate an FTB-3930 with

up to 10 FOT-930 units.

When measuring ORL (F

AS

T

ES

T or

ORL meter), accounts for

connector loss and adjusts ORL

calibration accordingly.
Not recommended for short links.

System

under test

Connector
mating

Connector

mating

System

under test

Connector
mating