beautypg.com

Measuring optical return loss, Connect a patchcord to the f – EXFO FTB-3930 MultiTest User Manual

Page 37

background image

Measuring Optical Return Loss

Multifunction Loss Tester

31

Performing ORL Reference and Setting ORL Zero Value

3. Connect a patchcord to the F

AS

T

ES

T SM port.

4. Terminate the fiber as close as possible before the component under

test. Wrap it at least 10 turns around a mandrel or small diameter tool,
adding turns until the reading stabilizes.

5. Press OK to save the ORL zero value, then remove the termination.

To revert to the factory-default ORL zero value:
1.
From the main window, select the ORL Meter tab.

2. Press Default ORL Zero.

Termination

Component
under test

FT

B

-3

930

MU

LT

IT

ES

T