13 Troubleshooting

Solving Common Problems

Problem

Possible Cause

Solution

During offset nulling, you

get the following message:
“Light detected during

nulling.”

Light reaches at least one

detector (power meter or
F

AS

T

ES

T).

Ensure protective caps are

tightly screwed on F

AS

T

ES

T

and power meter ports and

perform the nulling again.
Do not use rubber cover.

When using the power

meter, you get Power Too
Low
or Power Too High.

Power of the signal

received at the power

meter port is outside its

measurement range.

Check the connections.

Ensure you use the proper

fiber and connector type and

that you use your power

meter within specifications.

When measuring ORL, you

get Too Much Power.

Measured reflected power

is higher than emitted

power. The ORL reference

is incorrect.

Always perform ORL

reference before each ORL

measurement.

When measuring ORL, you

get ORL exceeds.

Reflection is below

sensitivity of ORL meter.

Perform an ORL zero

measurement to increase the

sensitivity of the detector.
Ensure mandrel is correct and

that patchcord and

connectors are in good

condition.