Performing Automated IL/ORL/Length Measurements (FasTesT)

36

FTB-3930

Setting Up the FasTest

Setting Up the FasTest

You can configure the F

AS

T

ES

T on the master unit only. The remote unit will

automatically adapt to these parameters (F

AS

T

ES

T parameters on remote

unit are ignored).

Setup includes the following elements:

Port and DUT (fiber) type: multimode F

AS

T

ES

T will not include ORL.

Compatibility: select FOT-930 / FTB-3930 for fast, two- or
three-wavelength testing including ORL (it requires two
FOT-930/FTB-3930). Use FOT-920 / FTB-3920 when other unit is an
FOT-920 or FTB-3920.

Mode/wavelengths: select one or more wavelengths for the F

AS

T

ES

T.

Depending on your choice, the F

AS

T

ES

T will include loss and/or ORL

measurements. Selecting FTTx Custom or FTTx All allows you to
define upstream and downstream wavelengths.

Unit location: in FTTx mode, you specify if the master unit is closer to
the CO (or to the premises) than the remote unit.