2200

ENMET Corporation

6


3.2 Sample Air Supply

Tap the pipe or tank containing the breathing air and use appropriate fittings to connect the sample input hose. The
instrument is designed to operate from an air supply pressure 55

PSIG

.

The sample air exits the instrument from the hole plug located on the bottom of the enclosure. Take care not to
obstruct this exit port.

3.3 Power Supply

The input power can vary from 100 to 240

V

AC

, 50/60 Hz. Mains power should be connected to the Power Input

Terminal

J23

and the ground screw

J21

. See Figure 2 for location.

W

ARNING

:

Continuous gas detection and alarm systems become inoperative upon loss of primary power.

Upon supplying air and power to the instrument:

The green power on LED is lit.

The display backlight is lit, and instrument will step through a start-up sequence: unit serial number, software

revision and gases monitored may be shown on the display.

The instrument may go into alarm briefly, but the sensors stabilize quickly. If the instrument persists in alarm,
acknowledge the alarm by pressing the

A

UDIO

D

EFEAT

/

A

LARM

A

CKNOWLEDGE

switch. If alarm persists longer than

30 minutes, call

ENMET

customer service personnel.

3.4 Outputs

Two types of alarm outputs are available, relay contacts and 4-20mA outputs.

Figure 2A: Relay, Input and Output Terminals

Relay 1
Channel 1
Alarm 1

Relay 2
Channel 2
Alarm 1

Relay 3
Channel 3
Alarm 1

Relay 4
Channel 4
Alarm 1

Relay 5
Channel 1-4
Alarm 2

Relay 6
System Fault

Connector

Dew Point Sensor

Connector

4-20mA Input

Connector

4-20mA Input

Connector 2
Channel 3 & 4
4-20mA Output

Connector 1
Channel 1 & 2
4-20mA Output

Connector RS485

Connector RS232

Connector RS485

Ground Screw J21

Power Input

Terminal J23

!