5.0 Maintenance

The

M

ED

A

IR

2200

requires periodic sensor calibration and replacement. Calibration of toxic gas and oxygen sensor

should be performed immediately following installation, one month after installation and every 3 months thereafter.
HC sensor should be calibrated on a monthly bases. Oxygen and CO sensor have an estimated lifetime of 1 – 3 years.
The CO

2

sensor has an estimated lifetime of 3-5 years. Sensors should be replaced when they will not calibrate or

shortly before the end of their estimated lifetime.

The dew point probe (sensor) can not be field calibrated. Dew point sensors should be replaced every two years.

5.1 Cleaning Instructions

C

AUTION

:

Never spray a cleaning solution on the surfaces of the

M

ED

A

IR

2200

devices.

Clean the exterior of the

M

ED

A

IR

2200

enclosures with a mild soap solution on a clean, damp cloth. Do not soak the

cloth with solution so that moisture drips onto, or lingers on, external surfaces.

Under no circumstances should organic solvents such as paint thinner be used to clean instrument surfaces.

5.2 Maintenance Menu

5.2.1 Accessing Maintenance Menu

The

M

ED

A

IR

2200

maintenance menu is accessed by entering the proper password with the

OPTION

and

SELECT

switches. See Section 5.2.2 Figure 6 for full Maintenance Menu flow chart.

Entrance to the maintenance menu is guarded with a four-digit Password. The factory default setting of the password
is 1270. When a valid numerical password is inserted, the user is allowed to enter the maintenance menu.

To enter the maintenance menu. Press the

OPTION

switch until “Enter Maint Menu” is displayed then press

SELECT

switch for the Enter Password menu. Enter the valid password as described below.

In the "

Enter Maint Menu

" position

Press the

SELECT

switch "

Enter Password

0

" is displayed. Press

SELECT

switch once, to move cursor to next

digit, this will be the first digit of the password.

In the

000

position, the underline cursor is under the left digit.

Press the

OPTION

switch to change the left digit; select the correct digit.

Press the

SELECT

switch, which locks the digit in place and moves the cursor one digit to the right.

Continue this process until the four-digit password is complete. When a valid password is inserted in this manner, the
display is transferred to the "

Calibration

" portion of the menu. If an invalid password is inserted you are returned to

the Enter Maint Menu display.

Example: Password Display (with factory installed password entered) and Flow Chart below.

To Calibration

See Section 5.2.2 Figure 6 for full Maintenance Menu flow chart.

Enter Password

1270

CO
PPM

O

2

%

O = Option Switch
S = Select Switch

DEW

POINT

O

S

S

Enter Password

0000

Enter Maint Menu

Changes digit indicated by underscore cursor

Locks underscored digit and moves cursor to next digit

O

(6)

Valid

Invalid

!