2.3 Circuit Board Features
The Display Panel is hinged on the left and is released by unscrewing the 2 screws located in the right corners. After
releasing the panel, it is swung to the left, exposing the interior of the enclosure. The Circuit Board is mounted at the
back surface of the enclosure interior. Features are shown in Figure 2.
Feature
Description
Relay Terminals
This group of terminals is located at the left side of the Circuit Board.
For the contacts for each of four alarm relays, and for the contacts of a fault relay.
Output Terminals
One 4-20mA output per active channel. 2 channels/outputs per connector.
Sensor Manifold
The sample manifold, the carbon monoxide, carbon dioxide and oxygen sensors are
located under this housing.
Dew Point Manifold
The dew point sensor is installed into this housing.
Figure 2: M
ED
A
IR
2200
Interior Features
Power Input
Terminal J23
Audio
Alarm
Sensor Dew Point
If Dew Point option is
installed
Relay Terminals
(6 places)
Aux Terminals
Optional (2 places)
Dew Point Terminal
Optional
Horn Terminal
Fuse Holders
0.630 Amp
(2 places)
Flow Sensor
Sensor
Manifold
Digital Communication
Terminals
Ground Screw
J21
Dew Point Manifold
If Dew Point option is
installed
4 – 20mA Output
Terminals
4 places