5.3.2 Set 4 –20mA Transmitter Scale

This section of the maintenance menu is installed when there are 4-20mA style sensors for dew point or other gases.

This function is normally performed at the factory and is not usually required to be performed in the field unless a
new transmitter is installed.

Consult the factory before entering this section of the maintenance menu.

After entering a valid password into maintenance menu, the Scale mA Xmtrs section is the second menu section, if it
is installed, enter by pressing the

S

ELECT

switch

Press the

SELECT

switch "

mA Xmter Scale: Select XX

" is displayed.

XX

= the gas to be set up.

Press the

OPTION

switch, if needed, to change to the gas to be set up.

Press the

SELECT

switch,

“Ch#: mAXmter: 4mA: 0000”

is displayed

Press the

SELECT

switch, that moves the cursor one digit to the right when the last digit is accepted the display

move to the full Scale mA Xmtrs menu

Press the

SELECT

switch,

“Ch#: mAXmter: 20mA: 0000”

is displayed

Press the

SELECT

switch, that moves the cursor one digit to the right when the last digit is accepted the display

will return to the Scale mA Xmtrs menu

Repeat these steps for each 4 –20mA transmitter.

Press

OPTION

switch until “Exit maint menu” appears and then press

SELECT

switch to return the instrument to the

Operational Display

Example: Sensor/Transmitter Set Up Flow Chart

✱1 N

OTE

: Press

OPTION

do not press

SELECT

, Consult the factory prior to selecting

SELECT

.

✱2 N

OTE

: Consult the factory before entering or making any adjustments.

O

1

S

S

2

mA Xmtr Scale

Select (Gas)

MAINTENANCE MENU

Scale mA Xmtrs

O

Pre ss

O

PTION

until the gas to be

S et Up is displayed

CH#: XX mA Xmter

4mA: 0000

S each digit

CH#: XX mA Xmter

20mA: 0000

S each digit

Example: Set up values

DP

-112F

68F

O = Press Option
S = Press Select