6.0 Technical Data and Specifications

Electrical Power

15 Amp fused branch circuit

100-240

V

AC

0.9 A

50/60 Hz

Board Mounted Fuse FH2, 0.630A, 5 x 20mm

Storage and Transport

Temperature:

-20

°

to +60

°

C (-4

°

to +140

°

F)

preferred

0

°

to +20

°

C (32

°

to 68

°

F)

Relative Humidity

0 - 99% RH, non-condensing

Atmospheric Pressure

20 to 36 inHg (68 to 133 kPa)

Operation

Temperature:

0

°

to +40

°

C (32

°

to +104

°

F)

Relative Humidity

0 - 99% RH, non-condensing

Atmospheric Pressure

20 to 36 inHg (68 to 133 kPa)

Air Line Pressure

55

PSI

(

± 5

PSI

)

Mechanical

Dimensions:

11 x 9 x 6 inches (4.3 x 3.5 x 2.4 cm)

Weight:

8 lbs (3.6 kg)

Material:

Engineered thermoplastic with hinged front cover

Strain relief:

5-12 mm OD

Outputs

Relays:

SPDT
Resistive Load

Inductive Load

10A at 110 VAC 7.5A at 110 VAC
10A at 30 VDC

5A at 30 VDC

Analog:

4-20 mA x 3

Digital:

RS-232 – Modbus
RS-485 – Modbus

Audio:

95 db at 2 ft

Sensors

Type

Range

Response Time

Life

CO

0 – 50 ppm

T

90

= 30 seconds

1 – 3 years

Dew Point

-112 - +68°F

T

90

= 10 seconds for –40°F to 50°F step change

2 years

O

2

0 – 30%

T

90

= 15 seconds

1 – 2 years

CO

2

0 – 5000 ppm

T

90

= 30 seconds

3 – 5 years

HC

0 – 100 ppm

T

90

= 30 seconds

1 – 2 years

N

OTE

:

All specifications stated in this manual may change without notice.