6.0 Technical Data and Specifications
Electrical Power
15 Amp fused branch circuit
100-240
V
AC
0.9 A
50/60 Hz
Board Mounted Fuse FH2, 0.630A, 5 x 20mm
Storage and Transport
Temperature:
-20
°
to +60
°
C (-4
°
to +140
°
F)
preferred
0
°
to +20
°
C (32
°
to 68
°
F)
Relative Humidity
0 - 99% RH, non-condensing
Atmospheric Pressure
20 to 36 inHg (68 to 133 kPa)
Operation
Temperature:
0
°
to +40
°
C (32
°
to +104
°
F)
Relative Humidity
0 - 99% RH, non-condensing
Atmospheric Pressure
20 to 36 inHg (68 to 133 kPa)
Air Line Pressure
55
PSI
(
± 5
PSI
)
Mechanical
Dimensions:
11 x 9 x 6 inches (4.3 x 3.5 x 2.4 cm)
Weight:
8 lbs (3.6 kg)
Material:
Engineered thermoplastic with hinged front cover
Strain relief:
5-12 mm OD
Outputs
Relays:
SPDT
Resistive Load
Inductive Load
10A at 110 VAC 7.5A at 110 VAC
10A at 30 VDC
5A at 30 VDC
Analog:
4-20 mA x 3
Digital:
RS-232 – Modbus
RS-485 – Modbus
Audio:
95 db at 2 ft
Sensors
Type
Range
Response Time
Life
CO
0 – 50 ppm
T
90
= 30 seconds
1 – 3 years
Dew Point
-112 - +68°F
T
90
= 10 seconds for –40°F to 50°F step change
2 years
O
2
0 – 30%
T
90
= 15 seconds
1 – 2 years
CO
2
0 – 5000 ppm
T
90
= 30 seconds
3 – 5 years
HC
0 – 100 ppm
T
90
= 30 seconds
1 – 2 years
N
OTE
:
All specifications stated in this manual may change without notice.