5.2.2 Maintenance Menu Flow Chart
The maintenance menu diagram is shown in Figure 6 Maintenance Menu Flow Chart. From the operational
display, press the
OPTION
switch 6 times; "
Enter
MAINTENANCE Menu
" is displayed.
Figure 6: M
ED
A
IR
2200 Maintenance Menu Flow Chart.
O = Press Option
S = Press Select
O
S
MAINTENANCE MENU
Exit maint menu
Press
O
PTION
to return to top of maintenance menu.
Press
S
ELECT
to return to operational menu.
See Section 5.3.9 for instructions on how to exit Maintenance Menus.
To/From
Operational
Display
From Operational Menu
Press O
PTION
(6 times)
S
Enter Maint Menu
O
S
Enter Password
0000
Changes digit indicated by underscore cursor
Locks underscored digit and moves cursor to next digit
O
(6)
Valid
Invalid
O
S
MAINTENANCE MENU
Set New Password
Password
XXXX
O
S
Changes digit indicated by underscore cursor
Locks underscored digit and moves cursor to next digit
See Section 5.3.8 for changing password instructions.
O
S
Calibration
Select (Gas)
MAINTENANCE MENU
Calibration
Press
O
PTION
until the channel to be Calibrated is displayed
See Section 5.3 for calibration instructions.
O
S
MAINTENANCE MENU
Set Alarm1
Alarm 1
Select: XX
Press
O
PTION
until the gas alarm to be Set is displayed
See Section 5.3.3 for setting alarms instructions.
O
S
Alarm 2
Select: XX
MAINTENANCE MENU
Set Alarm2
Press
O
PTION
until the gas alarm to be Set is displayed
See Section 5.3.3 for setting alarms instructions.
O
S
MAINTENANCE MENU
Set Alarm Delays
Alarm Delay
Select: XX
Press
O
PTION
until the gas alarm delay to be Set is displayed
See Section 5.3.4 for setting alarms instructions.
O
S
MAINTENANCE MENU
Configure Alarms
Ch 1, 2, 3, 4
R1 L
Pressing
O
PTION
changes letter indicated by underscore cursor
See Section 5.3.5 for configuring relay instructions.
O
S
MAINTENANCE MENU
Relay Failsafes
Relay Failsafes
R:1 Failsafe ON
Pressing
O
PTION
changes Failsafe setting from ON to OFF
See Section 5.3.6 for configuring relay failsafe instructions.
MAINTENANCE MENU
Scale mA Xmtrs
Press
O
PTION
until gas to be Set-Up is displayed
See Section 5.3.2 for transmitter set-up instructions.
If installed
Scale mA Xmtrs
Select (Gas)
O
S
O
S
MAINTENANCE MENU
mA Output Span
mA Output Span
Select: (Gas)
Press
O
PTION
until the gas span to be Set is displayed
See Section 5.3.7 for output span instructions.