6 operational menu, Figure 5: m, Operation menu flow chart – ENMET MEDAIR 2200 User Manual

Page 14

background image

M

ED

A

IR

2200

ENMET Corporation

11


4.6 Operational Menu

The operational menu allows the user to:

View alarm set point concentration values

View alarm ascending/descending trigger, latching and delay configurations

Enter the maintenance menu with the proper Password.

The operational menu is accessed with the

OPTION

and

SELECT

switches. The operational menu flow chart is shown in

Figure 5,

Pressing the

OPTION

switch is indicated with a "O"

Pressing the

SELECT

switch is indicated with a "S".

If the instrument is left at any location in the operational or maintenance menus, other than the operational display,
with no action taken for a period of 45 seconds, it returns to the operational display.

Figure 5: M

ED

A

IR

2200

Operation Menu Flow Chart

O

S

CH-1

CH-3

CH-2

CH-4

O = Press Option switch
S = Press Select switch

O

No
Function

Λ

L10 A1 vL19.5

vL-40

Λ

D 500

O

No
Function

S

S

Λ

L20 A2

Λ

23.5

Λ

L39

Λ

1000

O

No
Function

S

Enter Maint Menu

Enter Password

_

See Maintenance Menu Diagram

O

S

for each active channel

CH1 SCALE (CO)

0 – 50

PPM

S

CH3 SCALE

(O2)

0.0 – 30.0 %

CH2 SCALE (DP)

-112

– 68 F

S

CH4 SCALE (CO2)

0 – 5000

PPM

S

O

O

O

Displays are examples of gases:

Channel 1 = Carbon Monoxide

Channel 2 = Dew Point

Channel 3 = Oxygen

Channel 4 = Carbon Dioxide

✻ - Indicates relay is engaged
Displays are examples of Alarms

Λ

- Indicates alarm triggered on

increasing value of reading

v - Indicates alarm triggered on

decreasing value of reading

Displays are examples of Alarms

L

– Indicates alarm is in latching

mode.

(no

L

present) – Indicates alarm

is in non-latching mode.

Displays are examples of Alarms

D

– Indicates alarm is in

Differential Setting.

(no

D

present) – Indicates alarm

is in Standard Setting.

ALARM1 Delays
(Seconds)

O

S

No
Function

S

50 mASPAN 30.0
68

5000

O

No
Function

5

5

5

5

Alternating

S

Relays

1 2 3 4 5 6

✻=ON

✻✻✻✻✻✻

O

No
Function