ENMET Corporation

10


4.3 Alarm Latching or Differential Settings

An instrument is shipped with the alarms in the latching mode. The alarms may be independently configured in the
non-latching mode or differential setting by use of the maintenance menu.

See Section 5.3.3, for setting alarm 1 and alarm 2.

Standard Setting

I

N THE LATCHING MODE

: at the cessation of the condition which causes an alarm, the alarm indications do not

cease, and the alarm relay contacts do not revert to the non-alarm state, until the

AUDIO DEFEAT

/

ALARM

ACKNOWLEDGE

switch is pressed. An alarm can also be acknowledged by pressing the switch during the alarm

condition; then at the cessation of the alarm condition, alarm indications cease and alarm relays revert to the non-
alarm state. After an alarm is acknowledged, alarms in the latching configuration are re-armed to latch at the next
alarm condition.

I

N THE NON

-

LATCHING MODE

:

at the cessation of the condition that causes an alarm, the alarm indications

automatically cease, and the alarm relay contacts revert to the non-alarm state.

Differential Setting

The Alarm 1 differential value is the delay of the

M

ED

A

IR

2200

staying in alarm condition until after the measured

reading has returned past the alarm point by the differential value. Example: If the alarm point is

Λ

10 and the

differential is 2, the

M

ED

A

IR

2200

will go into alarm at 10 and stay in alarm until the reading has dropped below

8.

4.4 Audio Defeat

Pressing the

AUDIO DEFEAT

/

ALARM ACKNOWLEDGE

switch during an alarm temporarily silences the audio alarm.

Relays and alarm LEDs continue to function, in the alarm state, during an alarm condition. As long as the alarm
condition persists, the audio alarm will “chirp” every 20 seconds.

If after 15 minutes the alarm condition continues the audio alarm will reactivate at full intensity.

If any other alarm condition occurs while the audio alarm has been silenced it will force the audio alarm to

reactivate immediately.

4.5 Display

In clean air a display is shown in Figure 4

.

This position of the display is termed the "operational display". As

explained below, the display can be used to view other information by using the

OPTION

and

SELECT

switches.

Concentrations of CO and CO

2

are given in PPM (parts per million parts of air). Dew point is given in degrees

Fahrenheit at 55

PSIG

; this can be changed to degrees Centigrade by pressing the

SELECT

switch. Oxygen

concentration is given in percent by volume.

When sample flow is reduced below a limit, the bottom line of the display flashes “Low Flow Alarm”.