ENMET MEDAIR 2200 User Manual
Page 25
M
ED
A
IR
2200
ENMET Corporation
22
5.3.6 Failsafe Configuration
To change a relay failsafe configuration you must enter the maintenance menu. Press the
OPTION
switch until
“Enter Maint Menu” is displayed then press
SELECT
switch for the Enter Password menu. Enter the valid
password as described below.
In the "
Enter Maint Menu
" position
Press the
SELECT
switch "
Enter Password
❚
0
" is displayed. Press
SELECT
switch once, to move cursor to next
digit, this will be the first digit of the password.
In the
❚
000
position, the underline cursor is under the left digit.
Press the
OPTION
switch to change the left digit; select the correct digit.
Press the switch, which locks the digit in place and moves the cursor one digit to the right.
Continue this process until the four-digit password is complete. When a valid password is inserted in this manner,
the display is transferred to the "
Calibration
" portion of the menu. If an invalid password is inserted you are
returned to the Enter Maint Menu display.
After entering a valid password:
Press the
OPTION
switch until “Relay Failsafes” is displayed
Press the
SELECT
switch to indicate relay to be set.
Press the
OPTION
switch to set relay indicated, On or Off as appropriate.
Press the
SELECT
switch to cycle through each of the 5 relays, return to “Maintenance Menu Relay Failsafes”
Press
OPTION
switch until “Exit maint menu” appears and then press
SELECT
switch to return the instrument to the
Operational Display
Example: Set Relay Failsafe Configuration Flow Chart
5.3.7 Set Output Span Range
To change 4-20 mA output range. This range is set at the factory and should not be changed, contact
ENMET
for
information.
Press the
OPTION
switch to continue to next section of maintenance menu.
Press
OPTION
switch until “Exit maint menu” appears and then press
SELECT
switch to return the instrument to the
Operational Display
Example: Set Output Span Flow Chart
O = Press Option
S = Press Select
O
S
MAINTENANCE MENU
mA Output Span
S
O
S
mA OUTPUT SPAN
Changes
digit indicated by underscore cursor
Locks underscored digit and moves cursor to next digit
O
Press
O
PTION
until the channel to be Set is displayed.
Select: CO
mA OUTPUT SPAN
_ 50 CO
S
O = Press Option
S = Press Select
O
S
MAINTENANCE MENU
Relay Failsafes
S
Continue pressing
S
ELECT
to cycle through all 5 relays
Changing the setting of each relay as needed, using the
Option
switch
Pressing
O
PTION
changes Failsafe setting from ON to OFF
Relay Failsafes
R: 1 Failsafe ON
Relay Failsafes
R: 2 Failsafe ON
Relay Failsafes
R: 3 Failsafe ON
Relay Failsafes
R: 4 Failsafe ON
Relay Failsafes
R: 5 Failsafe ON