If system displays:
Fatal error - please restart the program
it means that an unrecoverable communication error has occurred during the cali
bration process(e.g. the monitor has been turned off).
Piease restart both the monitor and the calibration program.
If system displays:
Harnlog:
the
syst^
has
detected
a
configuration
error,
sor box does not match.........
The sen-
it means that the you have mismatched either the Light Probe unit or the base unit
of your caiibration system with a device belonging to a different 21XLP Light Probe
package.
The Light Probe and base unit combination within each calibration package are pre
cisely matched and tuned, for the maximum measurement accuracy.
Please check if both components of your calibration system are belonging to the
same 21XLP -2 Light Probe shipment.
If system displays:
Measure of light not correct
it means that 21XLP Light Probe zerois not properly placed on the screen.