If system displays:

Fatal error - please restart the program

it means that an unrecoverable communication error has occurred during the cali­
bration process(e.g. the monitor has been turned off).
Piease restart both the monitor and the calibration program.

If system displays:

Harnlog:

the

syst^

has

detected

a

configuration

error,

sor box does not match.........

The sen-

it means that the you have mismatched either the Light Probe unit or the base unit
of your caiibration system with a device belonging to a different 21XLP Light Probe
package.

The Light Probe and base unit combination within each calibration package are pre­
cisely matched and tuned, for the maximum measurement accuracy.

Please check if both components of your calibration system are belonging to the
same 21XLP -2 Light Probe shipment.

If system displays:

Measure of light not correct

it means that 21XLP Light Probe zerois not properly placed on the screen.