B Error messages

If system displays;

2lXliP not connected

It means that the application program is unable to locate the Light Probe unit at the

selected serial port. Check if aii the cabies are properiy connected and if the serial
port you specified is the right one.
Once you have fixed the problem, you need to switch the monitor off and then on
again to recover the error condition.

If system displays:

21XLP not connected

it means that the appiication program has detected a missing connection between

the base unit an the Light Probe unit of the 21XLP -2 Light Probe. Please check the

proper connection of the Light Probe cabie to the base unit.
Once you have fixed the probiem, you need to switch the monitor off and then on

again to recover the error condition.

if system displays:

Monitor not connected

it means that your dispiay does not respond to the caiibration program.
Please check if the sériai port of the monitor and the output port of the 21XLP -2
base unit are properiy connected.

if system displays:

2IXI1P setup error

means that during the Light Probe zero calibration (dark measure) an overfiow con­
dition occurred. Please assure that the Light Probe is properiy placed in its base
housing and restart calibration.