Operation – YSI ProDSS User Manual
Page 20
18
Operation
Figure 13 GLP Options
GLP Options
→ GLP → Options
User ID, Probe ID, or User Field #1 or 2 can be user-defined for positive
GLP file identification of:
• The person calibrating the instrument.
• The sensor/cable serial number used during calibration (or other,
user-defined Probe ID).
• Other user-specific identification (User Field #1 and #2) (
NOTE: User Field can be used to describe the condition of the
probe. For example, new sensor or new ODO cap.
Figure 14 Re-Cal Prompts
Re-Cal Prompts
→ GLP → Options → Re-Cal Prompts
Re-Cal Prompts provide a reminder to recalibrate a probe in the
user-defined number of days (
).
The Re-Cal prompt will be displayed in the message area of the main
display when the set time has elapsed (
).
Select the desired sensor Re-Cal prompt, then enter the desired number
of days before the Re-Cal prompt occurs.
Set the sensor value to zero (0) days (default) to turn off Re-Cal prompts.
NOTE: When enabled and the set amount of time since the last
calibration has passed, the Re-Cal prompt will be shown
when the instrument is turned on.
Figure 15 GLP Security
GLP Security
→ GLP → Security
The Calibration menu can be password protected to prevent accidental
or unauthorized sensor calibration (
).
1. From the GLP menu, select Security, then enter the default
password “ysi123”.
2. Select Set Password [] and change the default password.
3. Select the Protect Cal check box to password protect the
Calibration menu.
NOTE: Write down and keep the password in a safe place. Contact
YSI Technical Support if you lose the password (
).