3 ise measurement settings – YSI MultiLab 4010-2 User Manual
Page 88

Settings
MultiLab 4010-2
88
ba76142e03
08/2013
Default settings are printed in bold.
10.3 ISE measurement settings
The settings are made in the measuring menu of the ISE measurement. To
open the settings, activate the relevant measuring window in the measured
value display and press the
tings, switch to the measured value display with
The following settings are possible for ISE measurements:
Menu item
Possible setting
Explanation
mV resolution
0.1
1
Resolution of the mV display
Reset
-
Resets all sensor settings to the
delivery condition (see section
10.7.1 R
Menu item
Possible setting
Description
Calibration /
Calibration record
-
Displays the calibration
record of the last calibration.
Calibration /
Calibration data stor-
age /
Display
-
Displays the calibration
record.
Further options:
Scroll through the cali-
bration records with
<><>.
Output the displayed cal-
ibration record to the
interface with
[USB-Ausgabe].
Output all calibration
records to the interface
with
gabe].
Quit the display with
Switch directly to the
measured value display
with
Calibration /
Calibration data stor-
age /
Output to RS232/
USB
-
Outputs the calibration
records to the interface.