YSI MultiLab 4010-2 User Manual
Page 65

MultiLab 4010-2
Ion concentration
ba76142e03
08/2013
65
Measuring
7.
Start the measurement with
An entry window appears.
8.
Prepare the standard solution.
9.
If necessary, add the ISA/TISAB solution to the standard solution.
10.
Thoroughly rinse the electrode with deionized water.
11.
Immerse the electrode in the standard solution (complexing agent or
precipitating agent).
12.
Using <> <> and
standard solution (Std. volume), the concentration of the standard solu-
tion (Std. conc.) and the volume of the ISA/TISAB solution (ISA/TISAB
vol.).
13.
Select continue with <> <> and start the measurement with
When the measurement is finished an entry window appears.
14.
Add the sample to the standard (complexing agent or precipitating
agent).
15.
Using <> <> and
sample (Sample volume).
16.
Select continue with <> <> and start the measurement with
The measurement result is displayed when the measurement is com-
pleted.
i
Immerse sensor in standard
Std. volume
100.0 ml
Std. conc.
1.0 mg/l
ISA/TISAB vol.
1.0 ml
continue
Sample subtraction
01.08.2013 08:00
i
Add sample!
Sample volume
1.0 ml
continue
Sample subtraction
01.08.2013 08:00