ORP voltage

ba76142e03

08/2013

43

6

ORP voltage

6.1

Measuring

6.1.1 Measuring the ORP

Stability control

(AutoRead)

& HOLD function

The stability control function (AutoRead) continually checks the stability of the
measurement signal. The stability has a considerable impact on the reproduc-
ibility of measured values.
The measured parameter flashes on the display

as soon as the measured value is outside the stability range

when the automatic Stability control is switched off.

You can start the Stability control manually at any time, irrespective of the
setting for automatic Stability control (see section 10.6.3 A

UTOMATIC

S

TABILITY

CONTROL

, page 96) in the System menu.

The sensor connection and the USB-B (device) interface are gal-
vanically isolated. This facilitates interference-free measurements
also in the following cases:

Measurement in grounded test samples

Measurement with several sensors connected to one
MultiLab 4010-2 in one test sample

IDS ORP sensors are not calibrated. However, you can check IDS
ORP sensors using a test solution.

1.

Connect the ORP sensor to the meter.
The ORP measuring window is displayed.

2.

Adjust the temperature of the solutions and measure the current tem-
perature if the measurement is made without a temperature sensor.

3.

Check the meter with the ORP sensor.

4.

Immerse the ORP sensor in the test sample.