11 data storage, 1 manual data storage, 2 automatic data storage at intervals – YSI MultiLab 4010-2 User Manual
Page 101
MultiLab 4010-2
Data storage
ba76142e03
08/2013
101
11 Data storage
You can transmit measured values (datasets) to the data storage:
Manual data storage (see section 11.1 M
Automatic data storage at intervals (see section 11.2 A
11.1 Manual data storage
You can transmit a measurement dataset to the data storage as follows. The
dataset is at the same time output to the interface:
If the storage is full
When all storage locations are occupied, it is not possible to continue storing.
Then you can e.g. transmit the stored data to a PC or a USB flash drive (see
section 11.3.1 E
, page 104) and sub-
sequently erase the storage (see section 11.3.2 E
11.2 Automatic data storage at intervals
The storage interval (Interval) determines the time interval between automatic
data storage processes. Each data storage process transmits the current data-
set to the interface at the same time.
Configuring the
automatic storage
function
1.
Press the
The menu for manual data storage appears.
2.
If necessary, change and confirm the ID number (1 ... 10000) with
<><> and
The dataset is stored. The meter switches to the measured value dis-
play.
03.04.2013 11:24:16
pH 7.000 24.8 °C AR +++
ID number:
1
continue
Manual data storage 4 von 500
01.08.2013 08:00
1.
Press the
The menu for automatic data storage appears.