YSI MultiLab 4010-2 User Manual
Page 67

MultiLab 4010-2
Ion concentration
ba76142e03
08/2013
67
Measuring
01.08.2013 08:00
01.08.2013 08:00
#1 0.010 mg/l
20.2 mV 25.0 °C
#2 0.020 mg/l
79.2 mV 25.0 °C
Slp.: 54.1 mV
Sensor +++ (Ion type: Ag)
CALIBRATION
ADA 94/IDS BNC
B092500013
5.
Start the measurement with
An entry window appears.
6.
Prepare the sample.
7.
If necessary, add the ISA/TISAB solution to the sample.
8.
Thoroughly rinse the electrode with deionized water.
9.
Immerse the electrode in the sample that was supplemented with blank
value solution.
10.
Using <> <> and
sample (Sample volume), the volume of the ISA/TISAB solution (ISA/
TISAB vol.), the volume of the blank value solution (BV volume) and the
concentration of the blank value solution (BV conc.).
11.
Select continue with <> <> and start the measurement with
When the measurement is finished an entry window appears.
i
Immerse sensor in sample
Sample volume
100.0 ml
ISA/TISAB vol.
1.0 ml
BV volume
100.0 ml
BV conc.
1.000 mg/l
continue
Blank value addition
01.08.2013 08:00