2 measuring the temperature, Measuring the temperature – YSI MultiLab 4010-2 User Manual
Page 49
MultiLab 4010-2
Ion concentration
ba76142e03
08/2013
49
Criteria
The AutoRead criteria affect the reproducibility of the measured values. The
following criteria can be adjusted:
high: highest reproducibility
medium: medium reproducibility
low: lowest reproducibility
7.1.2 Measuring the temperature
For reproducible ion-selective measurements, it is essential to measure the
temperature of the test sample.
You have the following options to measure the temperature:
Measurement by an external temperature sensor.
Manual determination and input of the temperature.
The measuring instrument recognizes whether a suitable sensor is connected
and automatically switches on the temperature measurement.
The display of the temperature indicates the active temperature measuring
mode:
If you wish to measure (or calibrate) without temperature sensor, proceed as
follows:
3.
Using
or
Release the frozen measured value again with
The [AR] status display disappears. The display switches back to the
previous indication.
Increasing reproducibility also causes the response time to
increase until a measured value is evaluated as stable.
Temperatur
e sensor
Resolution of the
temp. display
Temp. measurement
yes
0.1 °C
Automatic with temperature sensor
-
1 °C
Manual
1.
Measure the current temperature of the test sample.
2.
Set the temperature value with <><>.
or
In the
with <><>.