MultiLab 4010-2

Ion concentration

ba76142e03

08/2013

49

Criteria

The AutoRead criteria affect the reproducibility of the measured values. The
following criteria can be adjusted:

high: highest reproducibility

medium: medium reproducibility

low: lowest reproducibility

7.1.2 Measuring the temperature

For reproducible ion-selective measurements, it is essential to measure the
temperature of the test sample.
You have the following options to measure the temperature:

Measurement by an external temperature sensor.

Manual determination and input of the temperature.

The measuring instrument recognizes whether a suitable sensor is connected
and automatically switches on the temperature measurement.

The display of the temperature indicates the active temperature measuring
mode:

If you wish to measure (or calibrate) without temperature sensor, proceed as
follows:

3.

Using , start a further measurement with stability control.
or
Release the frozen measured value again with or .
The [AR] status display disappears. The display switches back to the
previous indication.

Increasing reproducibility also causes the response time to
increase until a measured value is evaluated as stable.

Temperatur

e sensor

Resolution of the

temp. display

Temp. measurement

yes

0.1 °C

Automatic with temperature sensor

-

1 °C

Manual

1.

Measure the current temperature of the test sample.

2.

Set the temperature value with <><>.
or
In the /ISE/Man. temperature menu, set the temperature value
with <><>.