YSI MultiLab 4010-2 User Manual
Page 54
![background image](/manuals/716290/54/background.png)
Ion concentration
MultiLab 4010-2
54
ba76142e03
08/2013
Calibration
evaluation
After calibrating, the meter automatically evaluates the calibration.
Menu item
Setting/
function
Explanation
Calibration /
Calibration data stor-
age /Display
-
Displays the calibration records.
Further options:
Scroll through the calibration
records with <><>.
Using
played calibration record to the
USB-B (PC) interface or the USB-
A (USB printer) interface.
Using
tion records to the USB-B (PC)
interface or the USB-A (USB
printer) interface.
Quit the display with
Switch directly to the measured
value display with
Calibration / Calibra-
tion data storage /
Output to USB flash
drive or printer
-
Outputs the stored calibration data to
the USB-A interface (USB memory
device/USB printer)
Calibration /
Calibration data stor-
age /
Output to RS232/USB
-
Outputs the stored calibration data to
the USB-B interface (PC)
Display
Calibration
record
Magnitude of the slope [mV]
+++
50.0 ... 70.0 or 25.0 ... 35.0
Error
Error
< 50 or > 70
or
< 25 or > 35
Error elimination (see section 13 W
HAT
TO
DO
IF
..., page 89)