VACUUBRAND PC 600 LAN NT User Manual
Page 92
page 92 of 102
➨
Tighten the fillister head screws (L) of the
connection fasteners (Q) with a Torx driver
T20.
➨
Close the hinged covers.
Replace diaphragms and valves of the opposite side of
the pump in the same way.
➨
Feed the PTFE hose into the inlet of the
exhaust vapor condenser. Mount the con-
denser with the counter holder (Torx screws
T10) to the pump. Fasten union nut.
➨
Check the overpressure safety relief device
at the exhaust waste vapor condenser.
➨
If necessary, pull off the old pressure relief
valve and install the new one. Check for
correct position of the PTFE-foil under the
pressure relief valve.
➨
Assemble catchpots (9) with joint clips.
➨
Put head cover cowling on.
➨
Slide the head cover cowling in the grooves
of the caps (I) and under the connection
fasteners (Q).
➨
Install the washers. Use a Torx driver T20 to
attach the 4 screws holding the head cover
cowling.