Installing a pump in a vacuum system

If dangerous or polluting fluids could be released at

the outlet, install an appropriate system to catch and

dispose of those fluids.

+

Connect a gas-tight exhaust line at the pump outlet if

necessary. Always vent exhaust gases appropriately

(e.g., into a fume hood).

+

Never block the gas outlet. The exhaust line must al-

ways be free of obstructions (no back pressure) to en-

sure an unimpeded discharge of gas. The cross-sec-

tion of the outlet tubing must be at least the size of the

pump’s exhaust connection.

+

Maximum permissible pressure: 21.8 psi (1.5 bar) ab-

solute.

+

Attention: At pressures above approximately 795 Torr

(1060 mbar) the pressure reading becomes incorrect

due to saturation of the pressure transducer. The dis-

play will flash. Release pressure immediately! Risk of

bursting!

+

Particles and dust must not be aspirated. If necessary,

you must install appropriate filters. You must ensure

their suitability concerning gas flow, chemical resis-

tance and resistance to clogging prior to use.

Use and operation

When switching on the controller VNC 2 for the very first time, a menu

to select the language of the controller menu is displayed. Select the de-

sired language (e.g., ”English”) by pressing the t or s keys and press

”Enter” to confirm. Then select the pressure unit (”mbar”, ”Torr” or ”hPa”)

in the same way.

It is possible to access the language selection menu at any time via

the Device menu.