Technical data, Protection, “technical data”, pg. 25) – VACUUBRAND PC 600 LAN NT User Manual
Page 25
![background image](/manuals/715222/25/background.png)
page 25 of 102
Technical data
Type
PC 500 LAN NT PC 600 LAN NT
Maximum pumping speed*
50/60 Hz (ISO 21360)
cfm
(m
3
/h)
1.2 / 1.4
(2.0 / 2.3)
2.0 / 2.2
(3.4 / 3.8)
Ultimate vacuum (absolute)
without gas ballast
Torr
(mbar)
5.3
(7)
1.1
(1.5)
Ultimate vacuum (absolute) with
gas ballast
Torr
(mbar)
9.0
(12)
2.3
(3)
Maximum permissible inlet
pressure (absolute)
psi
(bar)
16
(1.1)
Maximum permissible outlet
pressure (absolute)
psi
(bar)
16
(1.1)
Maximum pressure
difference between inlet and
outlet
psi
(bar)
16
(1.1)
Maximum permissible pressure
(absolute) at gas ballast valve
psi
(bar)
17.5
(1.2)
Permissible ambient
temperature
storage / operation
°F
(°C)
14 to 140 / 50 to 104
(-10 to +60 / +10 to +40)
Permissible relative atmospheric
moisture during operation (no
condensation)
%
30 to 85
Maximum permissible installation
altitude above mean sea level
ft
(m)
6500
(2000)
Rated motor power
hp
(kW)
0.24
(0.18)
0.34
(0.25)
No-load speed 50/60 Hz
rpm
1500 / 1800
Maximum permissible range
of supply voltage ( ±10% )
Attention: Observe specifications
of rating plate!
100-115 V~ 50/60 Hz, 120 V~ 60 Hz
230 V~ 50/60 Hz
Maximum rated current at:
100-115 V~ 50/60 Hz,
120 V~ 60 Hz
230 V~ 50/60 Hz
A
A
3.4
1.8
5.7
3.0
Device fuse
2 slow blow fuses 6.3A
Motor protection
thermal cutout, manual reset
* Pumping speed of diaphragm pump