beautypg.com

Gas inlet temperatures, Abbreviations, Pump parts – VACUUBRAND PC 600 LAN NT User Manual

Page 29

background image

page 29 of 102

Abbreviations

ETFE: Ethylene/Tetrafluoroethylene

ECTFE: Ethylene/Chlorotrifluoroethylene

FFKM: Perfluoro elastomer

FPM: Fluoroelastomer

PET: Polyethylene terephthalate

PPS: Polyphenylene sulfide

PTFE: Polytetrafluoroethylene

Position

Component

1

Inlet

2

Outlet

3

Gas ballast valve

4

ON/OFF switch

5

Mains connection

6

Handle

7

Pump rating plate

8

Controller VNC 2

Position

Component

9

Catchpot

10

Exhaust waste vapor

condenser

11

Overpressure safety relief

device

12

Coolant inlet

13

Coolant outlet

14

Fan

Pump parts

We reserve the right for technical modification without prior notice!

Gas inlet temperatures

* if pumping potentially explosive atmospheres: 50 °F to 104 °F (+10°C to +40°C)

Operating condition

Inlet pressure

Permitted range of gas

temperatures at inlet

Continuous operation

> 75 Torr (100 mbar)

(high gas load)

50 °F to 104 °F

(+10°C to +40°C)

Continuous operation

< 75 Torr (100 mbar)

(low gas load)

32 °F to 140 °F*

(0°C to +60°C*)

Short-time

(< 5 minutes)

< 75 Torr (100 mbar)

(low gas load)

14 °F to 176 °F*

(-10°C to +80°C*)

This manual is related to the following products: