USL LSS-100 User Manual
Page 18
LSS−100P Installation & Operation Manual
18
lss100.sys.capture_cct|Color Temperature|6100|6300
6262
Instructs the LSS-100P to capture
the correlated color temperature
in Kelvin to its log. The
parameters are description, min,
and max
lss100.sys.capture_temperature_c|Temperature|20|40 25
Instructs the LSS-100P to capture
its internal temperature in
Celcius to its log. The parameters
are description, min, and max.
lss100.sys.capture_temperature_f|Temperature|60|80 70
Instructs the LSS-100P to capture
its internal temperature in
Fahrenheit to its log. The
parameters are description, min,
and max.
lss100.sys.capture_x|Color|0.308|0.320
0.314
Instructs the LSS-100P to capture
the color coordinate x. The
parameters are description,
Xmin, Xmax.
lss100.sys.capture_y|Color|0.345|0.357
0.351
Instructs the LSS-100P to capture
the color coordinate y. The
parameters are description,
Ymin, Ymax.
lss100.sys.log_post|http://lss.uslinc.com/xml/post.php
Posts logs in XML to the specified
URL. The URL can be HTTP or
HTTPS. No port numbers are
allowed. Posts all log data since
the last log post to this URL. If
more than one post is included
in a script, add a wait of 60
seconds between posts to allow
time for the first to complete.
lss100.sys.script_run|0
Runs the specified script. If no
script number is specified, runs
script 0 (the first one).
lss100.sys.script_wait|30
Causes a script to wait the
specified number of seconds
before executing the next line.
Waits between captures keeps
the measurement capture
synchronized with the test
content.
Figure 19 - LSS-100P TCP Commands