LSS−100P Installation & Operation Manual
17
The LSS-100P commands and typical responses are shown in figure 19. Note that many of these
commands are commonly used in scripts (described previously), and the lss100.sys.script_run command
can be used to start script execution.
Command
Response
Description
lss100.sys.auditorium
1
Returns the auditorium number
entered on the configuration
page.
lss100.sys.cct
6255.250284
Returns the correlated color
temperature.
lss100.sys.cdm2
405.151492
Returns the luminance in
candelas per square meter
lss100.sys.comments
Conference Room
Returns the auditorium
comments entered on the
configuration page.
lss100.sys.ftl
118.332596
Returns the measured luminance
in foot-Lamberts
lss100.sys.spl
83.239520
Returns the measured sound
pressure level in dBc
lss100.sys.temperature
29.2\t84.
Returns the internal
temperature in degrees C and
degrees F (separated by tab).
lss100.sys.theater_name
USL San Luis Obispo Returns the theater name
entered on the configuration
page.
lss100.sys.theater_number
1
Returns the theater number
entered on the configuration
page.
lss100.sys.ver
C\t130522\t120530
Returns the circuit board
version, the firmware version,
and the bootloader versions,
separated by tab.
lss100.sys.xy
0.314743\t0.353419 Returns the chromaticity as x tab
y.
lss100.sys.capture_spl|SPL Center Low|81.2|83.2
83.239520
Instructs the LSS-100P to capture
the current SPL to its log. The
parameters are description, min,
and max. Returns the current
SPL in dBc.
lss100.sys.capture_fl|Luminance|13|15
14.1234
Instructs the LSS-100P to capture
the luminance (measured in fL)
to its log. The parameters are
description, min, and max.
lss100.sys.capture_cdm2|Luminance|47|49
48.4321
Instructs the LSS-100P to capture
the luminance (measured in
cd/m
2
) to its log. The parameters
are description, min, and max.