beautypg.com

0 maintenance and service, 1 media pads – Reznor REC (Evaporative Cooling) Unit Installation Manual User Manual

Page 9

background image

Form I-REC, P/N 160201 R3, Page 9

FIGURE 12A -

Adjust Water Flow

with the Ball Valve

in FIGURE 12B.

FIGURE 12B -

AquaSaver Water

Line showing

Solenoid Valve

and Ball Valve

Solenoid

Valve

Ball

Valve

A

Sprinkler Pipe

Pad Height A = Water rise from PVC Sprinkler Pipe

24” (610mm) 1/8” to 1/2” (3 to 13mm)
48”(1219mm) 1/4” to 1/2” (6 to 13mm)

2) AquaSaver Timer Adjustment - At any given tem-

perature, the media pads should completely wet from top

to bottom during the ON cycle. The microprocessor has

three preset timing settings based on media size. The

appropriate setting is selected by changing the position of

the suitcase jumper at J2 on the microprocessor. Remove

the cover and check the setting (See

FIGURE 13).

If the jumper is at the appropriate location for the media

(“S” for 24” or “M” for 48”), replace the cover. If the jumper

needs to be moved, move it to the appropriate setting.

The setting will go into effect when the power is restored.

Check the “ON” timing; the media pads should be wet

from top to bottom during the ON cycle.
If the preset timing is not suitable for the application, fol-

low the instructions supplied with the microprocessor to

change the calibration of the “ON” and/or “OFF” cycle.

NOTE: Prior to 2003 the AquaSaver timed cycle was controlled

by a mechanical timer. Turn the adjustment screw clockwise to

increase the ON time or counterclockwise to decrease the ON
time. One complete turn will adjust the cycle by 12-14 seconds.

FIGURE 13 -

AquaSaver

Microprocessor

Control in the

Junction Box

J2

1 2 3 4

1 2 3 4

1 2 3 4

1 2 3 4

S

M

L

12-36”

(305-914mm)

Media

Height

37-48”

(940-1219mm)

Media

Height

49-72”

(1245-1829mm)

Media

Height

6.0 Maintenance and Service

WARNING:

Disconnect all

power to the unit

before doing any

maintenance. Failure

to do so may cause

electrical shock,

personal injury or

death.

6.1 Media Pads

Over time, excessive amounts of mineral deposits will begin to build up on the media.

Annually, scale and dirt should be washed off the entering surface of the media.

Remove the pad retainers and screen. Clean the media using a garden hose, mild

soap, and a soft bristled brush. When the media becomes too clogged with mineral

deposits and dirt that it cannot be cleaned, the pads should be replaced. The average

pad life is approximately three cooling seasons.
Select the correct replacement part numbers and order replacement media pads from

your Reznor Distributor. Follow the instructions on page 10 to remove and replace

pads as shown in

FIGURE 14

REC Media Pad

Size

Qty

Replacement P/N

12” Cellulose

12” Glass Fiber

40

24x12

2

106021

106029

24x2-3/8

1

106022

106030

50

24x12

2

106021

106029

24x7-7/8

1

106023

106031

60

24x12

3

106021

106029

24x1-3/8

1

106024

106032

70

24x12

3

106021

106029

24x9-5/8

1

106025

106033

80

24x12

4

106021

106029

24x2-7/8

1

106026

106034

90

24x12

4

106021

106029

24x8-5/8

1

106027

106035

180

48x12

4

107194

107201

48x8-5/8

1

107195

107202

360

48x12

8

107194

107201

48x8-5/8

2

107195

107202

Media Pads

Screen

Screen Retainer

Top Pad

Retainer

Screw

Bottom Pad

Retainer

FIGURE 14 - Removal

and Replacement of

Evaporative Cooling

Media