beautypg.com

Specifications product specifications, Environmental conditions, Electrical specifications – Oakworks DTPM300 User Manual

Page 19

background image

16

SPeCIfICATIOnS

PRODUCT SPeCIfICATIOnS

Materials of Construction

This product contains no latex.

Top Frame

extruded aluminum with powdercoated or anodized surfaces.

Table Top

.5” (1.3 cm.) Polycarbonate

Patient Comfort

1” (2.5 cm.) upholstered foam pad

Table Length

78” (198 cm.)

Table Width

24” (61 cm.)

Height range

24”-40” (61-102 cm.) standard

Table Top Travel range

±11.5” (29 cm)

Trendelenburg Tilt

±14°

Weight

390 lbs. (177 kg.)

shipping Weight

526 lbs. (239 kg.)

Lifting Capacity

450 lbs. (205 kg.)

Aluminum Filtration

equivalent of Table top

TAbLe TOP

100 kvp, Hvl of 3.6 mm = 1.62 (1.74 max)

1” (2.5 cm.) TAbLe TOP PAD

100 kvp, Hvl of 3.6 mm = .88 (1.07 max)

2” (5 cm.) TAbLe TOP PAD

100 kvp, Hvl of 3.6 mm = 1.47 (1.76 max)

This product complies with the United States Department of Health and Human Services radiation performance

standards, 21 CFR Subchapter I, in effect at the time of manufacture for radiographic tables.

envIROnMenTAl COnDITIOnS

Conditions

Temperature

Humidity

Atmospheric Pressure

Normal Use

50° (10°C) to 104° (40°C)

20% to 60% rH

98 to 105 kPa

storage & Transport

-20° (-29°C) to 135° (57°C)

20% to 95% rH

98 to 105 kPa

eleCTRICAl SPeCIfICATIOnS

Designed for:

North America

europe

Japan

Input service

120 VAC/15 amp/60 Hz

220 VAC/10 amp/50/60 Hz

100 VAC/20 amp/50 Hz

Current Draw

5.8 amps

2.9 amps

4.8 amps

Maximum Momentary Cur-

rent Consumption

6.0 amps

3.0 amps

7.2 amps

Voltage Output to Actuators

24 VDC

24 VDC

24 VDC

electric shock Protection

Class 1 equipment

Class 1 equipment

Class 1 equipment

Tabletop Applied Part

Type b Applied Part

Type b Applied Part

Type b Applied Part

Ingress Protection rating

IPX0

IPX0

IPX0

Made of Operation

Intermittent Operation

MAX 2 minutes ON

18 minutes off

Intermittent Operation

MAX 2 minutes ON

18 minutes off

Intermittent Operation

MAX 2 minutes ON

18 minutes off