10

DIReCTIOnS fOR USe

CARbOn fIbeR ARM bOARD (ACCeSSORy)

CAUTIOn

Do not place undue weight or downward pressure on the Carbon Fiber Arm Board. It is a position-

ing device for the arms and should not be used as leverage to get on or off the table. Injury can

occur. DO NOT exceed 30 lbs. (13.6 kg.) of load on the

Support Section. Injury can occur.

Use

This radiolucent device can be used with any flat support surface where ionizing radiation (x-ray) is utilized for imag-

ing. It can be used to support arms and feet.

Directions

Unfold the Arm Board and slide the Base Section under the patient and the table pad. Position the patient and the

Support Section

to suit the needs of the procedure. The weight of the patient will hold the Arm Board in place.

Rotate the Support Section to the desired position. This section rotates up to 180º.

Prone Position at 135º

Supine Position at 180º

Support Section

Base Section

To remove the Arm Board, ask the patient to move their limb off the Support Section and lift their body slightly. This

will allow you to pull out the base section of the Arm Board.
Fold the Base and Support sections for storage.

flUORO exTenDeR (ACCeSSORy)

CAUTIOn

Do not place undue weight or downward pressure on the Fluoro Extender. It is a positioning device

for the arms and should not be used as leverage to get on or off the table. Injury can occur.

The Oakworks

®

Fluoro Extender is radiolucent and offers extra width only where you need it for

greater positioning accuracy with optimum imaging. The Fluoro Extender is inserted under the table

top pad.