KERN & SOHN TN 300-0.01US User Manual
Instruction manual tn_us, Sauter gmbh
Sauter GmbH
Tieringerstr. 11-15
D-72336 Balingen
E-Mail: [email protected]
Tel: +49-[0]7433- 9976-174
Fax: +49-[0]7433-9976-285
Internet: www. sauter.eu
Instruction Manual
TN_US
TN_US-BA-e-1112
1
DIGITAL ULTRASONIC THICKNESS GAUGE
Models: TN 80-0.1US
TN 230-0.1US
TN 300-0.1US
TN 80-0.01US
TN 230-0.01US
TN 300-0.01US
Table of contents
1. Overview
1.1 Product specifications
1.2 Main functions
1.3 Measuring principle
1.4 Configuration
1.5 Operating conditions
2. Structure feature
2.1 Measurement screen
2.2 Keypad definition
3. Preparation
3.1 Transducer selection
3.2 Conditions and preparation of surfaces
4. Operation
4.1 Power on/ off
4.2 Zero adjustment
4.3 Sound velocity calibration
4.4 How to perform measurements
4.5 Scan Mode
4.6 Changing resolution
4.7 Changing units
4.8 Memory management
4.9 Data printing
4.10 “Beep”- Mode
4.11 EL Backlight
4.12 Battery information
4.13 Auto Power Off
4.14 System reset
4.15 Connection to PC
5. Servicing
6. Transport and storage
Appendix A Sound velocities
Appendix B Application notes
7. Declaration of conformity
1. Overview
Model TN- US is a digital ultrasonic thickness gauge
based on the same operating principles as SONAR. The
instruments are capable of measuring the thickness of
various materials with an accuracy of 0.1/0.01 mm. They
are suitable for a variety of metallic and non- metallic
materials.
1.1 Product specifications
Display: 4.5 digits LCD with EL backlight
Measuring range: 0.75 to 300mm (in steel)
Sound velocity: 1000 to 9999m/s
Resolution: TN xx0.1 US: 0,1mm;
TN xx0.01US: 0,1 / 0,01mm
- Model TN 80-0.01measures continuously
with a resolution of 0.01
- Model TN 230-0.01 US as well as
TN 300-0.01 are measuring with a
resolution of 0.01 up to 200mm
and over this, each device measures with a
resolution of 0.1
Accuracy: Models with a resolution of 0.1mm:
0.5% of the measured value +0.04mm
Models with a resolution of 0.01mm:
1% of the measured value
In dependence on material and environmental
conditions.
Units: Metric/ Imperial units selectable
-
Four measurements readings per second at
single point measurement and ten per second at
Scan Mode.
-
Memory up to 20 files (up to 99 values for each
file) of stored values
Power supply: 2x AA, 1.5V alkaline batteries
Typical operating time: about 100 hours
(EL backlight off)
Transfer to PC: RS-232 serial port for TN xx0.01 US.
No transfer to PC possible at TN xx0.1 US
Dimensions: 150 x 74 x 32 mm
Weight: 245g
1.2 Main functions
-
capable of performing measurements on a wide
range of materials including metals, plastic,